Flow-injection spectrophotometric determination of copper (II) ion at low μg L-1 levels


Asan A.

International Journal of Environmental Analytical Chemistry, cilt.79, sa.2, ss.111-119, 2001 (SCI-Expanded) identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 79 Sayı: 2
  • Basım Tarihi: 2001
  • Doi Numarası: 10.1080/03067310108035903
  • Dergi Adı: International Journal of Environmental Analytical Chemistry
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.111-119
  • Anahtar Kelimeler: Copper(II) determination, Flow-injection, N,N′-o-phenylenebis(salicylaldimine), Spectrophotometry
  • Ondokuz Mayıs Üniversitesi Adresli: Evet

Özet

A highly selective, sensitive and fast determination of trace amounts of copper (II) ion is described. The method is based on the formation of the coloured Cu(II)-N,N′-o-phenylenebis(salicylaldimine) (PBS) complex, which is then selectively monitored at 415 nm. To increase the interaction between copper(II) and the colour forming reagent and preconcentrate copper(II), a microcolumn containing strong cation-exchange resins was placed between the injection manifold and the spectrophotometer. The linear range of the proposed method was 25-100 μg L-1, and the limit of detection (signal-to-noise ratio = 2) was < 1 μg L-1 for 20 μL injection volume of copper (II) ion solution at a sampling rate of > 50 h-1. The relative standard deviation at the level of 25-100 μg L-1was 0.24 % for n = 5. The precision and accuracy of the method was checked by analysis of a certified reference material, and the applied to fiver and sea water samples.